Proline Promag 55 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 131
692 S: SIM. MEASURAND
!: # 692
0x60
0x61
0x62
UNCERTAIN
UNCERTAIN
Simulated
Value
(manually
specified
value)
O.K.
Low
High
Simulation
measured value
active
Cause:
Simulation of the measured value is active.
Remedy:
Switch off simulation.
Access:
SUPERVISION SYSTEM OPERATION
SIM. MEASURAND ( OFF)
698 S: DEV. TEST ACT.
!: # 698
0x60
0x61
0x62
UNCERTAIN
UNCERTAIN
Simulated
Value
(manually
specified
value)
O.K.
Low
High
Device test via
Fieldcheck active
Cause:
The measuring device is being checked on site via
the test and simulation device.
No. # 8xx Software option errors
840 S: COAT.E1DEVLIM
! # 840
0x40
0x41
0x42
UNCERTAIN
UNCERTAIN
Non specific
(uncertain
status)
O.K.
Low
High
Delay time outside
the limit
Cause:
The measured error of the decay time constant is
outside the range defined in the WARNING (7546)
function.
Remedy:
Remove the sensor from the pipe and check
whether the inside wall of the measuring tube has
to be cleaned.
841 S: COAT.E2DEVLIM
!: # 841
0x40
0x41
0x42
UNCERTAIN
UNCERTAIN
Non specific
(uncertain
status)
O.K.
Low
High
Delay time outside
the limit
Cause:
The measured error of the decay time constant is
outside the range defined in the WARNING (7546)
function.
Remedy:
Remove the sensor from the pipe and check
whether the inside wall of the measuring tube has
to be cleaned.
845 S: COATING FAILURE
!: # 845
0x40
0x41
0x42
UNCERTAIN
UNCERTAIN
Non specific
(uncertain
status)
O.K.
Low
High
Coating detection
not possible
Cause:
Coating detection not possible:
1. The recovery time entered for this is too small.
2. The measuring tube is empty or not filled
completely.
Remedy:
1. Increase the value for the recovery time (
RECOVERY TIME function, 7523).
2. Fill the measuring tube (if necessary, check the
process conditions of the facility).
846 S: NOI.VAL.DEVLIM
!: # 846
0x40
0x41
0x42
UNCERTAIN
UNCERTAIN
Non specific
(uncertain
status)
O.K.
Low
High
Noise valve outside
the limit
Cause:
The measured error of the decay time constant is
outside the range defined in the WARNING (7546)
function.
Remedy:
Check the application for process change (pressure,
air bubbles, inhomogenity etc.).
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts ä 135 ff.)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits