Installation Proline Promag 55 PROFIBUS DP/PA
14 Endress+Hauser
Partially filled pipes
Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection
function offers additional protection by detecting empty or partially filled pipes ä 117.
"
Caution!
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable
to install a cleaning valve.
A0003204
Fig. 8: Installation in partially filled pipe
Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes longer than 5 meters (16 ft).
This precaution is to avoid low pressure and the consequent risk of damage to the lining of the
measuring tube. This measure also prevents the system losing prime, which could cause air
inclusions.
Information on the lining's pressure tightness ä 148
A0011902
Fig. 9: Measures for installation in a down pipe (h > 5 m / 16 ft)
1 Vent valve
2Siphon