Proline Promag 55 PROFIBUS DP/PA Commissioning
Endress+Hauser 75
!
Note!
• The display returns to the cell SETUP COMMISSIONING (1002) if you press the ESC key combination during
parameter interrogation. The stored parameters remain valid.
• The "Commissioning" Quick Setup must be carried out before any of the other Quick Setups described in these
Operating Instructions are run.
m Only units not yet configured in the current setup are offered for selection in each cycle. The unit for mass and
volume is derived from the corresponding flow unit.
n The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
o This prompt only appears if a current output and/or pulse/frequency output is available. Only the outputs not yet
configured in the current Setup are offered for selection in each cycle.
p The "YES" option remains visible until all the outputs have been configured.
"NO" is the only option displayed when no further outputs are available.
q The "automatic display configuration" option contains the following basic settings/factory settings
YES Main line = Volume Flow
Additional line = Totalizer 1
Information line = Operating/System Conditions
NO The existing (selected) settings remain.
➅ The BATCHING QUICK SETUP is only available when the optional software package BATCHING is installed.
➆ The PULSATING FLOW QUICK SETUP is only available if the device has a current output or a pulse/frequency
output.