Proline Promag 55 PROFIBUS DP/PA Commissioning
Endress+Hauser 75
!
Note!
The display returns to the cell SETUP COMMISSIONING (1002) if you press the ESC key combination during
parameter interrogation. The stored parameters remain valid.
The "Commissioning" Quick Setup must be carried out before any of the other Quick Setups described in these
Operating Instructions are run.
m Only units not yet configured in the current setup are offered for selection in each cycle. The unit for mass and
volume is derived from the corresponding flow unit.
n The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
o This prompt only appears if a current output and/or pulse/frequency output is available. Only the outputs not yet
configured in the current Setup are offered for selection in each cycle.
p The "YES" option remains visible until all the outputs have been configured.
"NO" is the only option displayed when no further outputs are available.
q The "automatic display configuration" option contains the following basic settings/factory settings
YES Main line = Volume Flow
Additional line = Totalizer 1
Information line = Operating/System Conditions
NO The existing (selected) settings remain.
The BATCHING QUICK SETUP is only available when the optional software package BATCHING is installed.
The PULSATING FLOW QUICK SETUP is only available if the device has a current output or a pulse/frequency
output.