Commissioning Proline Promag 55 PROFIBUS DP/PA
80 Endress+Hauser
6141 SET UNIT TO BUS If this function is executed, the measured variables are transmitted
cyclically to the PROFIBUS master (Class 1) with the system units
set in the measuring device.
Options:
OFF
SET UNITS (transmission is started by pressing F key)
"
Caution!
Activating this function can cause a sudden change of the
measured variables transmitted to the PROFIBUS master (Class 1);
this, in turn, can affect subsequent control routines.
Æ
Back to the HOME position:
Press and hold down Esc keys X for longer than three seconds or
Repeatedly press and release Esc keys X Exit the function matrix step by step
Quick Setup "Communication"