Proline Promass E 300 PROFINET Diagnostics and troubleshooting
Endress+Hauser 165
Diagnostic information Remedy instructions
No. Short text
046 Sensor limit exceeded 1. Inspect sensor
2. Check process condition
Measured variable status [from the factory]
1)
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal S
Diagnostic behavior Warning
Influenced measured variables
Oscillation amplitude 1
Oscillation amplitude 2
Application specific output
Application specific output
Signal asymmetry
Carrier mass flow
Carrier pipe temperature
Target corrected volume flow
Carrier corrected volume flow
Concentration
Oscillation damping 1
Oscillation damping 2
Density
Oil density
Water density
Dynamic viscosity
Sensor electronic temperature (ISEM)
GSV flow
GSV flow alternative
Kinematic viscosity
Mass flow
Oil mass flow
Water mass flow
Index inhomogeneous medium
Index suspended bubbles
HBSI
NSV flow
NSV flow alternative
External pressure
Exciter current 1
Exciter current 2
Oscillation frequency 1
Oscillation frequency 2
S&W volume flow
Reference density
Reference density alternative
Corrected volume flow
Oil corrected volume flow
Water corrected volume flow
Oscillation damping fluctuation 1
Oscillation damping fluctuation 2
Frequency fluctuation 1
Frequency fluctuation 2
Target mass flow
Carrier volume flow
Target volume flow
Temp. compensated dynamic viscosity
Temp. compensated kinematic viscosity
Temperature
Status
Volume flow
Oil volume flow
Water volume flow
Water cut
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.