Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
199
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
276 I/O module 1 to n faulty 1. Restart device
2. Change I/O module
0x100007B
0x1000081
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
283 Memory content 1. Reset device
2. Contact service
0x10000E1
0x100016F
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
302 Device verification active Device verification active, please
wait.
0x20001EE
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
311 Electronic failure 1. Do not reset device
2. Contact service
0x40000E2
Status signal M
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
332 Writing in HistoROM backup failed Replace user interface board
Ex d/XP: replace transmitter
0x10002C7
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
361 I/O module 1 to n faulty 1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
0x1000095
Status signal F
Diagnostic behavior Alarm