Proline Promass F 300 EtherNet/IP System integration
87
Possible input values
Possible input values 1 to 10:
Off
Mass flow
Volume flow
Corrected volume flow
Target mass flow
1)
Carrier mass flow
1)
Target volume flow
1)
Carrier volume flow
1)
Target corrected volume flow
1)
Carrier corrected volume flow
1)
Density
Reference density
Concentration
1)
Temperature
Carrier tube temperature
2)
Electronic temperature
Oscillation frequency 0
Oscillation frequency 1
2)
Oscillation amplitude 0
Oscillation amplitude 1
2)
Frequency fluctuation 0
Frequency fluctuation 1
2)
Oscillation damping 0
Oscillation damping 1
Tube damping fluctuation 0
Tube damping fluctuation 1
Signal asymmetry
Exciter current 0
Exciter current 1
2)
Monitoring of exciter current 0
Monitoring of exciter current 1
2)
HBSI
2)
Totalizer 1
Totalizer 2
Totalizer 3
Alternative reference density
3)
GSV flow
3)
Alternative GSV flow
3)
NSV flow
3)
Alternative NSV flow
3)
S&W volume flow
3)
1) Only available with the Concentration application package.
2) Only available with the Heartbeat Verification application package.
3) Only available with the Petroleum application package.
Possible input values 11 to 20:
Off
Actual diagnostics
Previous diagnosis
Mass flow unit
Volume flow unit
Corrected volume flow unit
Temperature unit
Density unit
Reference density unit
Concentration unit
Current unit
Totalizer 1 unit
Totalizer 2 unit
Totalizer 3 unit
Verification results
Verification status
Status zero point adjustment
Permanently assigned output group
Output assembly fix (Assem102), 54 byte
Description (format) Byte Bit Value
1. Totalizer 1 1 0
0: Enable
1: Disable
2. Totalizer 2 1
3. Totalizer 3 2
4. Verification 3
5. Concentration of medium type 4
6. Compensation, pressure 5
7. Reference density compensation 6
8. Compensation, temperature 7
9. S&W correction value % 2 0
0: Enable
1: Disable
10. Water cut % 1
11. Flow override 2
12. Zero point adjustment 3
13. Not used 4
14. Not used 5
15. Not used 6
16. Not used 7