6 Commissioning PROline Prosonic Flow 93 PROFIBUS-DP/-PA
102 Endress+Hauser
Input
byte
Process parameter Access
type
Comment/Data format Factory
setting
unit
0, 1, 2, 3
Volume flow CH1
(Channel: 273)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m
3
/h
4
Status
volume flow CH1
read Status code
Page 112
5, 6, 7, 8
Sound velocity CH1
(Channel: 293)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m/s
9
Status
sound velocity CH1
read Status code
Page 112
10, 11, 12, 13
Flow velocity CH1
(Channel: 309)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m/s
14
Status
flow velocity CH1
read Status code
Page 112
15, 16, 17, 18
Volume flow CH2
(Channel: 529)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m
3
/h
19
Status
volume flow CH2
read Status code
Page 112
20, 21, 22, 23
Sound velocity CH2
(Channel: 549)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m/s
24
Status
sound velocity CH2
read Status code
Page 112
25, 26, 27, 28
Flow velocity CH2
(Channel: 565)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m/s
29
Status
flow velocity CH2
read Status code
Page 112
30, 31, 32, 33
Average volume flow
(Channel: 567)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m
3
/h
34
Status
average volume flow
read Status code
Page 112
35, 36, 37, 38
Average sound velocity
(Channel: 570)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
m/s
39
Status
average sound velocity
read Status code
Page 112
Total volume flow
(Channel: 568)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
can be assigned as required!
m
3
/h
Status
total volume flow
read Status code
Page 112
Volume flow difference
(Channel: 569)
read
32-bit floating point number
(IEEE-754)
Illustrated
Page 99
can be assigned as required!
m
3
/h

Table of Contents