PROline Prosonic Flow 93 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 127
Compatibility
Amp. - I/O Mod.
S A / C COMPATIB.
$ # 051
51 device failure BAD
0x0F
constant
Cause:
The I/O board and the amplifier
board are not compatible.
Remedy:
Use only compatible assemblies
and boards.
Check the compatibility of the
assemblies used. Check the:
– Spare part set number
– Hardware revision code
F-CHIP failure S HW F-CHIP
$ # 061
61 device failure BAD
0x0F
constant
Cause:
Transmitter F-Chip
1. F-Chip is defective.
2. F-Chip is not plugged into the
I/O board or is missing.
Remedy:
1. Replace the F-Chip.
Accessories
Page 136
2. Plug the F-Chip into the I/O
board
Page 138
Interruption between
sensor and transmitter
CH1
S SENS. DOWN CH1
$ # 082
82 sensor failure BAD
0x13
constant
Cause:
System error. Connection between
sensor channel 1/2 and transmitter
interrupted.
Remedy:
– Check the cable connection
between the sensor and the
transmitter.
– Check that the sensor connector
is fully screwed in.
– The sensor may be defective.
– Incorrect sensor connected
Change sensor type specification:
• PROFIBUS (Commuwin II):
Manufacturer-specific
Transducer Block
Sensor data
(V3H1)
• Local display:
BASIC FUNCTIONS
SENSOR
DATA
SENSOR PARAMETER
SENSOR TYPE
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy