PROline Prosonic Flow 93 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 127
Compatibility
Amp. - I/O Mod.
S A / C COMPATIB.
$ # 051
51 device failure BAD
0x0F
constant
Cause:
The I/O board and the amplifier
board are not compatible.
Remedy:
Use only compatible assemblies
and boards.
Check the compatibility of the
assemblies used. Check the:
Spare part set number
Hardware revision code
F-CHIP failure S HW F-CHIP
$ # 061
61 device failure BAD
0x0F
constant
Cause:
Transmitter F-Chip
1. F-Chip is defective.
2. F-Chip is not plugged into the
I/O board or is missing.
Remedy:
1. Replace the F-Chip.
Accessories
Page 136
2. Plug the F-Chip into the I/O
board
Page 138
Interruption between
sensor and transmitter
CH1
S SENS. DOWN CH1
$ # 082
82 sensor failure BAD
0x13
constant
Cause:
System error. Connection between
sensor channel 1/2 and transmitter
interrupted.
Remedy:
Check the cable connection
between the sensor and the
transmitter.
Check that the sensor connector
is fully screwed in.
The sensor may be defective.
Incorrect sensor connected
Change sensor type specification:
PROFIBUS (Commuwin II):
Manufacturer-specific
Transducer Block
Sensor data
(V3H1)
Local display:
BASIC FUNCTIONS
SENSOR
DATA
SENSOR PARAMETER
SENSOR TYPE
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy