9 Trouble-shooting PROline Prosonic Flow 93 PROFIBUS-DP/-PA
128 Endress+Hauser
Interruption between
sensor and transmitter
CH2
S SENS. DOWN CH2
$ # 083
83 sensor failure BAD
0x13
constant
Cause:
System error. Connection between
sensor channel 1/2 and transmitter
interrupted.
Remedy:
Check the cable connection
between the sensor and the
transmitter.
Check that the sensor connector
is fully screwed in.
The sensor may be defective.
Incorrect sensor connected
Change sensor type specification:
PROFIBUS (Commuwin II):
Manufacturer-specific
Transducer Block
Sensor data
(V3H1)
Local display:
BASIC FUNCTIONS
SENSOR
DATA
SENSOR PARAMETER
SENSOR TYPE
Interruption between
sensor and transmitter
CH1
S SENSOR UP CH1
$ # 085
85 sensor failure BAD
0x13
constant
Cause:
System error. The connection
between sensor and transmitter is
interrupted.
Remedy:
Check the cable connection
between the sensor and the
transmitter.
Check that the sensor connector
is fully screwed in.
The sensor may be defective.
Incorrect sensor connected
Change sensor type specification:
PROFIBUS (Commuwin II):
Manufacturer-specific
Transducer Block
Sensor data
(V3H1)
Local display:
BASIC FUNCTIONS
SENSOR
DATA
SENSOR PARAMETER
SENSOR TYPE
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy