PROline Prosonic Flow 93 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 129
Interruption between
sensor and transmitter
CH2
S SENSOR UP CH2
$ # 086
86 sensor failure BAD
0x13
constant
Cause:
System error. Connection between
sensor channel 1/2 and transmitter
interrupted.
Remedy:
Check the cable connection
between the sensor and the
transmitter.
Check that the sensor connector
is fully screwed in.
The sensor may be defective.
Incorrect sensor connected
Change sensor type specification:
PROFIBUS (Commuwin II):
Manufacturer-specific
Transducer Block
Sensor data
(V3H1)
Local display:
BASIC FUNCTIONS
SENSOR
DATA
SENSOR PARAMETER
SENSOR TYPE
TOT could not be
restarted
S CHECKSUM. TOTAL.
$ # 111
111 device failure BAD
0x0F
constant
Cause:
System error. Totalizer checksum
error
Remedy:
1. Restart the measuring device
2. Replace the amplifier board if
necessary.
Spare parts
Page 136
SW. Compatibility
AMP-I/O Modul
S A / C COMPATIB.
$ # 121
121 device failure BAD
0x0F
constant
Cause:
System error. Due to different
software versions, I/O board and
amplifier board are only partially
compatible (possibly restricted
functionality).
Remedy:
Replace module with lower
software version.
Spare parts
Page 136
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy