PROline Prosonic Flow 93 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 129
Interruption between
sensor and transmitter
CH2
S SENSOR UP CH2
$ # 086
86 sensor failure BAD
0x13
constant
Cause:
System error. Connection between
sensor channel 1/2 and transmitter
interrupted.
Remedy:
– Check the cable connection
between the sensor and the
transmitter.
– Check that the sensor connector
is fully screwed in.
– The sensor may be defective.
– Incorrect sensor connected
Change sensor type specification:
• PROFIBUS (Commuwin II):
Manufacturer-specific
Transducer Block
Sensor data
(V3H1)
• Local display:
BASIC FUNCTIONS
SENSOR
DATA
SENSOR PARAMETER
SENSOR TYPE
TOT could not be
restarted
S CHECKSUM. TOTAL.
$ # 111
111 device failure BAD
0x0F
constant
Cause:
System error. Totalizer checksum
error
Remedy:
1. Restart the measuring device
2. Replace the amplifier board if
necessary.
Spare parts
Page 136
SW. Compatibility
AMP-I/O Modul
S A / C COMPATIB.
$ # 121
121 device failure BAD
0x0F
constant
Cause:
System error. Due to different
software versions, I/O board and
amplifier board are only partially
compatible (possibly restricted
functionality).
Remedy:
Replace module with lower
software version.
Spare parts
Page 136
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy