9 Trouble-shooting PROline Prosonic Flow 93 PROFIBUS-DP/-PA
130 Endress+Hauser
Save to T-DAT failed S LOAD T-DAT
$ # 205
205 device failure BAD
0x0F
constant
Cause:
System error. Data backup to T-DAT
failed or error when accessing the
calibration values stored in the
T-DAT.
Remedy:
1. Check whether the T-DAT is
correctly plugged into the
amplifier board
Page 138
2. Replace the T-DAT if it is
defective. Spare parts
Page 136
Check the spare part set
number to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
3. Replace measuring electronics
board if necessary.
Spare parts
Page 136
Restore from T-DAT
failed
S SAVE T-DAT
$ # 206
206 device failure BAD
0x0F
constant
Cause:
System error. Data backup to T-DAT
failed or error when accessing the
calibration values stored in the
T-DAT.
Remedy:
1. Check whether the T-DAT is
correctly plugged into the
amplifier board
Page 138
2. Replace the T-DAT if it is
defective. Spare parts
Page 136
Check the spare part set
number to ensure that the new
replacement DAT is compatible
with the measuring electronics.
3. Replace measuring electronics
board if necessary.
Spare parts
Page 136
Communication failure S COMMUNICAT. I/O
$ # 261
261 no communication BAD
0x18
no limits
Cause:
System error. Communication error.
No data reception between
amplifier and I/O board or faulty
internal data transfer.
Remedy:
Check whether the electronics
boards are correctly inserted in
their holders.
Page 138
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy