Troubleshooting • 9
1
Troubleshooting
CHAPTER 9
Error message Possible cause What to do ?
I 10 Underdose 1. Blocked hose
2. Cylinder empty
3. System error
1. Remove tissue from instrument
2. Change cylinder
3. Technical service
I 11 Overdose 1. Input pressure too high
2. System error
1. Replace pressure reducer
2. Technical service
I 12 Overdose 1. Input pressure too high
2. System error
1. Replace pressure reducer
2. Technical service
I. 13 Input pressure 1. No cylinder connected
2. System error
1. Connect cylinder
2. Technical service
I 150 Button pressed 1. When switching on, a button on
front panel was pressed
2. Button on front panel faulty
1. Do not press any button on the
front panel
2. Technical service
I 160 CUT-foot switch
pressed
1. When switching on, CUT-foot
switch pressed
2. Foot switch faulty
1. Do not press any foot switch
when switching on
2. Technical service
I 161 COAG-foot switch
pressed
1. When switching on, COAG-foot
switch pressed
2. Foot switch faulty
1. Do not press any foot switch
when switching on
2. Technical service
165 CUT-finger switch
pressed
1. When switching on, CUT-finger
switch pressed
2. Finger switch faulty
1. Do no press any finger switch,
when switching on
2. Technical service
I 166 COAG-finger
switch pressed
1. When switching on, COAG-finger
switch pressed
2. Finger switch faulty
1. Do no press any finger switch,
when switching on
2. Technical service
I 200 Programming error
system page
Error reported after a page was
programmed
Technical service
I 201 System page not
initialized
System parameters incomplete Technical service
I 250 Blocked hose Tissue in instrument Remove tissue from instrument
I 251 Pressure limitation
I 262 Cylinders 1 and 2
are empty
Replace cylinders
I 263 Cylinder 1 is
empty, 2 is nearly empty
Replace cylinder 1
I 264 Cylinder 2 is
empty, 1 is nearly empty
Replace cylinder 2
I 265 Cylinders 1 and 2
are nearly empty
Use up one cylinder and then
replace