S2000 Elite - User's Guide
035-00707 Revision 1
36
While the CAL button on the radiometer is pressed, the display will indicate the current setpoint
and the SET icon will be flashing. If the CAL button on the radiometer is pressed for less than
5 seconds, the current radiometer set-point will be sent to the S2000 Elite.
Holding this button for 5 seconds will store the current optical input into the radiometer’s set
point (this feature can be enabled or disabled via PC). The SET Icon will cease flashing and
be continuously on until the button is released.
When the CAL button is pressed for less than 5 seconds and released, the radiometer will
send the setpoint to the S2000 Elite, thus setting the S2000 Elite in absolute
mode. Additionally, the current light guide diameter value selected by the radiometer will be
transferred to the S2000 Elite. This light guide diameter will be used by the S2000 Elite to
calculate the proper output irradiance in W/cm
2
or power in W
Whenever the calibration procedure is activated, the screen on the S2000 Elite will change to
the screen below.
For detailed operating instructions regarding calibration of the S2000 Elite, refer to
the R2000 user’s guide.