Operate the unit in a well ventilated area with at least six inches clearance at the rear
of the unit for proper air flow. Do not place any objects below the unit, between the
feet as this will restrict airflow through the bottom of the front face plate.
It is recommended that the unit to be situated on a benchtop, shelf or rack when in
use.
For safe operation, use only a grounded outlet.
Avoid physical shocks or jarring to the unit especially while the unit is operating. Such
sudden movements reduce the lamp module life.
The lamp module must be operated for a minimum of 20 minutes each time it is
turned on to prevent damaging the lamp. Increasing the time between turning the
lamp module on and off will maximize lamp life.
Replace the air filter, found under the front faceplate, frequently to ensure
unrestricted airflow. It is recommended that the air filter be removed and washed with
a mild detergent and water every time the lamp module is replaced.
Restricted airflow can cause the lamp temperature to increase above optimum
temperature, significantly reducing lamp life.
1. When necessary, clean the light-emitting end of the light guide using an optical cleaning
solution.
2. Cleaning of the unit is not required; however, if cleaning is desired, disconnect the AC
power cord from the unit and use de-ionised water, isopropyl alcohol or an ammonia-
based glass cleaning fluid. Ensure that the cleaning solution does not come in contact
with any optical, moving mechanical or electrical parts.
3. Recommended operation of S2000 Elite is in a horizontal position. Using S2000 Elite
4. beyond 15% of tilt from a flat position will wear lamp life.
• Replacing the External Fuses
• Replacing the Air Filter
• Light Guide Cleaning