A Menu structure FLUXUS F808, F809
UMFLUXUS_F808_8091V1-2-1EN, 2019-08-23 171
activation of the input of a lower limit of the flow velocity:
• factory: the default limit of 2.5 cm/s will be used
• user: input of a limit
x
input of the cut-off flow for positive measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user has been selected.
x
input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user has been selected.
x
input of the cut-off flow for the absolute value of the measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = absolut and Cut-off
Flow = user has been selected.
x
input of the max. signal amplification
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
x
input of the fixed upper limit of the sound speed
range: 0...3 000 m/s
0: the default value 1 848 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the offset
range: 0...900 m/s
0: the default value 300 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
activation of the overflow of the totalizers x
activation of the taking-over of the totalizer values after a restart of the measure-
ment
x
input of the time interval without any valid measured values after which the
transmitter recognizes a long measurement failure
0: the default value 30 s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the number of decimal places for the totalizers
Automatic: dynamic adjustment
Fixed to x digit: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
x
activation of the manual reset of the totalizers.
This display will only be indicated if the SuperUser mode is activated.
x
INIT-
resistant
Cut-off Flow
factory >USER<
+Cut-off Flow
2.5 cm/s
-Cut-off Flow
-2.5 cm/s
Cut-off Flow
2.5 cm/s
A:Gain threshold
Fail if > 90 dB
A:Bad soundspeed
thresh. 2007 m/s
A:Bad soundspeed
offset: +321 m/s
Quant. wrapping
off >ON<
Quantity recall
off >ON<
Do not total. if
no meas.> 0 s
Total digits ↕
Automatic
3xC clear totals
off >ON<