A Menu structure FLUXUS F808, F809
UMFLUXUS_F808_8091V1-2-1EN, 2019-08-23 171
activation of the input of a lower limit of the flow velocity:
factory: the default limit of 2.5 cm/s will be used
user: input of a limit
x
input of the cut-off flow for positive measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user has been selected.
x
input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user has been selected.
x
input of the cut-off flow for the absolute value of the measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = absolut and Cut-off
Flow = user has been selected.
x
input of the max. signal amplification
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
x
input of the fixed upper limit of the sound speed
range: 0...3 000 m/s
0: the default value 1 848 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the offset
range: 0...900 m/s
0: the default value 300 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
activation of the overflow of the totalizers x
activation of the taking-over of the totalizer values after a restart of the measure-
ment
x
input of the time interval without any valid measured values after which the
transmitter recognizes a long measurement failure
0: the default value 30 s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the number of decimal places for the totalizers
Automatic: dynamic adjustment
Fixed to x digit: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
x
activation of the manual reset of the totalizers.
This display will only be indicated if the SuperUser mode is activated.
x
INIT-
resistant
Cut-off Flow
factory >USER<
+Cut-off Flow
2.5 cm/s
-Cut-off Flow
-2.5 cm/s
Cut-off Flow
2.5 cm/s
A:Gain threshold
Fail if > 90 dB
A:Bad soundspeed
thresh. 2007 m/s
A:Bad soundspeed
offset: +321 m/s
Quant. wrapping
off >ON<
Quantity recall
off >ON<
Do not total. if
no meas.> 0 s
Total digits
Automatic
3xC clear totals
off >ON<

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F808 A1 Series and is the answer not in the manual?

Flexim FLUXUS F808 A1 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F808 A1 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals