CommentsValueLabel
Move the joystick left/right to enable or disable
noise reduction.
When Noise reduction is set to On, the image
noise decreases and the image appears more
stable.
However, when the camera or the object moves,
and Noise reduction set to On, this may create
some image smearing.
On
Off
Noise reduction
Press the Adjust region button to display a region
on the screen that will be used when autoadjusting
the camera.
Adjust region
Press the joystick left/right to change the shutter
period, or switch off the shutter.
Please note the following:
Although the shutter period works independent-
ly of other functions described in this publica-
tion, FLIR Systems recommends that Short is
selected when using the camera for detection
of face temperature.
Selecting Normal will calibrate the camera at
least every 15th minute, while selecting Short
will calibrate the camera at least every 3rd
minute.
If the shutter is switched off, a symbol (*) will
prefix the result at the time a shutter sequence
should have taken place, thus indicating uncer-
tainty in the measurement result.
Normal
Short
Off
Shutter period
10.2.5.2 Difference
10393203;a3
Figure 10.41 Difference settings dialog box
Difference is a command that calculates the temperature difference between two
measurement markers, or the reference temperature and a measurement marker.
10
92 Publ. No. 1557966 Rev. a155 ENGLISH (EN) February 6, 2006
10 Camera program