Chapter 4: Maintenance
Nebulizer Assembly Maintenance
CyTOF XT User Guide 115
NOTE The larger slit opening of the nebulizer mount should align with the connection
for the nebulizer line on the spring block. The pin on the spring block fits into the
smaller slit on the nebulizer mount.
10 Insert the spring assembly into the nebulizer mount and secure it by engaging both pins
to lock it in place. Place your thumb on the flat surface of the spring assembly and
turning clockwise to lock the pins in place.
11 Refer to Reassemble the Nebulizer Assembly to prepare the nebulizer for immediate
use or storage.