Chapter 4: Maintenance
Clean the Vacuum Interface Cones
CyTOF XT User Guide 69
8 Remove the skimmer-reducer assembly using the cone removal tool. Insert the other
end of the cone removal tool (with 2 magnets) into the 2 holes on the skimmer-reducer
assembly.
9 Turn the cone removal tool counterclockwise to unscrew the skimmer-reducer assembly
from the vacuum interface.