5502A
Operators Manual
4-26
Caution
To prevent damage to the UUT, makes sure the applied voltage to
the UUT is not larger than the rating of the UUT insulation and the
connection wires.
1. Push to clear any output from the Calibrator.
2. Connect the UUT as in the “Connect the Calibrator to a UUT” section of this chapter.
Change the voltage and current connections to this procedure.
3. Set the UUT to measure dc power on the necessary range.
4. Push the numeric keys and decimal point key to record the necessary voltage output
(maximum seven numeric keys). For example, 123.4567.
Note
At voltage outputs of 100 V and above (nominal), the Calibrator usually
makes a small high-pitched sound.
5. Push to select the polarity of the voltage (default is +).
6. Push a multiplier key, if necessary. For example, push .
7. Push .
8. The Control Display shows the amplitude of your entry. For example, 123.4567 mV
(see below).
nn071f.eps
9. Push the numeric keys and decimal point key to record the necessary current output
(maximum six numeric keys). For example, 234.567.
10. Push to select the polarity of the current (default is +).
11. Push a multiplier key, if necessary. For example, push .
12. Push .
13. The Control Display shows the amplitude of your entries. For example, 123.4567 mV
and 234.567 mA ( see below).
nn082f.eps