5502A
Operators Manual
1-4
waves to 2 MHz.
A Standard IEEE-488 (GPIB) interface that complies with ANSI/IEEE Standards
488.1-1987 and 488.2-1987.
A EIA Standard RS-232 serial data interface to print, show, or move internally stored
calibration constants, and for remote control of the 5502A.
A pass-through RS-232 serial data interface to send data to the Unit Under Test
(UUT).
Safety Information
In this manual, A Warning identifies conditions and procedures that are dangerous to the
user. A
Caution identifies conditions and procedures that can cause damage to the
Product or the equipment under test.
Warning
To prevent possible electrical shock, fire, or personal injury:
Use the Product only as specified, or the protection
supplied by the Product can be compromised.
Carefully read all instructions.
Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
Use this Product indoors only.
Do not touch voltages > 30 V ac rms, 42 V ac peak, or
60 V dc.
Do not use the Product if it operates incorrectly.
Do not use and disable the Product if it is damaged.
Do not use test leads if they are damaged. Examine the test
leads for damaged insulation, exposed metal, or if the wear
indicator shows. Check test lead continuity.
Use only cables with correct voltage ratings.
Connect the common test lead before the live test lead and
remove the live test lead before the common test lead.
Use only the mains power cord and connector approved for
the voltage and plug configuration in your country and rated
for the Product.
Make sure the ground conductor in the mains power cord is
connected to a protective earth ground. Disruption of the
protective earth could put voltage on the chassis that could
cause death.