EasyManua.ls Logo

Fluke 5502A - AC Current (Non-Sine Wave) (Cont.); AC Current, Square Wave Characteristics (Typical); AC Current, Triangle Wave Characteristics (Typical)

Fluke 5502A
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5502A
Operators Manual
1-28
AC Current (Non-Sine Wave) (cont.)
Square Wave Range p-p Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
0.047 to 0.65999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
0.66 to 6.59999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
6.6 to 65.9999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
66 to 659.999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.5
1 to 10 kHz 10 + 2
0.66 to 5.99999 A
[2]
10 to 45 Hz 0.5 + 1.0
45 Hz to 1 kHz 0.5 + 0.5
1 to 10 kHz 10 + 2
6 to 41 A
[2]
45 to 500 Hz 0.5 + 0.5
500 Hz to 1 kHz 1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
AC Current, Square Wave Characteristics (typical)
Range LCOMP Risetime Settling Time Overshoot
I <6 A @ 400 Hz off
25 μs 40 μs to 1 % of final value
<10 % for <1 V Compliance
3 A & 20 A Ranges on
100 μs 200 μs to 1 % of final value
<10 % for <1 V Compliance
AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range

Table of Contents

Other manuals for Fluke 5502A

Related product manuals