5502A
Operators Manual
1-12
Resistance
Range
[1]
Absolute Uncertainty, tcal ±5 °C ±(% of output + floor)
[2]
Resolution
(Ω)
Allowable Current
[3]
(A)
% of output
Floor (Ω) Time and temp since ohms
zero cal
90 Day 1 Year
12 hrs ±1 °C 7 days ±5 °C
0 to 10.999 Ω
0.009
0.012
0.001 0. 01 0.001 1 mA to 125 mA
11 to 32.999 Ω
0.009
0.012
0.0015 0.015 0.001 1 mA to 125 mA
33 to
109.999 Ω
0.007
0.009
0.0014 0.015 0.001 1 mA to 70 mA
110 to
329.999 Ω
0.007
0.009
0.002 0.02 0.001 1 mA to 40 mA
330 to
1.09999 kΩ
0.007
0.009
0.002 0.02 0.01 1 mA to 18 mA
1.1 to
3.29999 kΩ
0.007
0.009
0.02 0.2 0.01
100 μA to 5 mA
3.3 to
10.9999 kΩ
0.007
0.009
0.02 0.1 0.1
100 μA to 1.8 mA
11 to
32.9999 kΩ
0.007
0.009
0.2 1 0.1
10 μA to .5 mA
33 to
109.999 kΩ
0.008
0.011
0.2 1 1
10 μA to 0.18 mA
110 to
329.999 kΩ
0.009
0.012
2 10 1
1 μA to 50 μA
330 kΩ to
1.09999 MΩ
0.011
0.015
2 10 10
1 μA to 18 μA
1.1 to
3.29999 MΩ
0.011
0.015
30 150 10
250 nA to 5 μA
3.3 to
10.9999 MΩ
0.045
0.06
50 250 100
250 nA to 1.8 μA
11 to
32.9999 MΩ
0.075
0.1
2500 2500 100 25 nA to 500 nA
33 to
109.999 MΩ
0.4
0.5
3000 3000 1000 25 nA to 180 nA
110 to
329.999 MΩ
0.4
0.5
100000 100000 1000 2.5 nA to 50 nA
330 to
1100.00 MΩ
1.2 1.5 500000 500000 10000 1 nA to 13 nA
[1] Continuously variable from 0 Ω to 1.1 GΩ.
[2] Applies for 4-WIRE compensation only. For 2-WIRE and 2-WIRE COMP, add 5 μV per amp of stimulus current to the floor
specification. For example, in 2-WIRE mode, at 1 kΩ the floor specification within 12 hours of an ohms zero cal for a measurement
current of 1 mA is: 0.002 Ω + 5 μV / 1 mA = (0.002 + 0.005) Ω = 0.007 Ω.
[3] Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases by Floor
(new)
= Floor
(old)
x
I
min
/I
actual
. For example, a 50 μA stimulus measuring 100 Ω has a floor specification of: 0.0014 Ω x 1 mA/50 μA = 0.028 Ω, assuming
an ohms zero calibration within 12 hours.