EasyManua.ls Logo

Fluke 5502A - Verify RS-232 Host Port Operation with a Terminal

Fluke 5502A
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5502A
Operators Manual
5-12
Verify the RS-232 Host Port
Choose or adapt one of the subsequent procedures to verify the Calibrator RS-232 Host
port connected to a PC COM port. A typical connection is shown in Figure 5-4. Note that
a null modem cable is used for connection. (See Appendix C for information about
RS-232 cables and connectors.)
UUT
SERIAL 1
FROM HOST
Port
5502A Calibrator
Controller
COM Port
Null Modem Cable
gjh044.eps
Figure 5-4. RS-232 Host Port Verification
Terminal This procedure uses the Terminal accessory supplied with Windows (or equal)
to verify RS-232 Host port operation. To use this method, you must select term as the
Remote I/F in Step 4 in the procedure “RS-232 Host Port Setup Procedure.”
Visual Basic This procedure uses Visual Basic (see Appendix D) to verify RS-232 Host
port and RS-232 UUT port operation.
Verify RS-232 Host Port Operation with a Terminal
Complete the subsequent procedure to verify RS-232 Host port operation with the
Windows Terminal accessory (or equal).
1. Complete the “RS-232 Host Port Setup Procedure” before in this chapter to set up the
Calibrator for RS-232 Host port operation. Write down the RS-232 Host port
parameters that you selected in this procedure.
2. Connect the selected COM port on the PC to the Calibrator SERIAL 1 FROM HOST
port with a standard null-modem RS-232 cable. (See Appendix C for information on
RS-232 cables and connectors.)
3. Open Windows to the Program Manager screen on your PC.

Table of Contents

Other manuals for Fluke 5502A

Related product manuals