5502A
Operators Manual
10-10
2. To move the cursor between the voltage and frequency fields, push , see below.
gl004i.eps
3. Use the and keys to move the cursor to the digit that you will change.
4. Turn the rotary knob to change the value.
When you use the rotary knob in volt mode or marker mode, the Control Display
shows the percentage change of the new value from the reference value. This is
useful to learn the percentage of error on the oscilloscope. You can set the reference
value to the new value. Push ,see below.
gl005i.eps
5. Push to remove the cursor from the Output Display and save the new value as
the reference value.
Note
If you try to use the rotary knob to adjust a value to an quantity that is
invalid for the function that you use, or is out of the range limit of the value,
the value will not change and the 5502A will beep. If it is necessary to go to
a different range of values, turn the knob quickly to jump to the new range.
Use and
The and keys cause the current value of the signal to jump to a predetermined
cardinal value. The current function determines this value. These keys are explained more
under the descriptions for each function.
Reset the Oscilloscope Option
You can reset all parameters in the 5502A to their default settings during front panel
operations. Push on the front panel.
After you reset the 5502A, push to go to the Oscilloscope Calibration Option (the
Volt menu shows). Push to connect the signal output.