D-1
Appendix D
Error Messages
Error Messages
The subsequent is a list of the Calibrator error messages. The error message format is
shown in Table D-1.
Table D-1. Error Messages Format
Error Number (Message Class : Description) Text Characters
0 to 65535 QYE Query Error, caused by a full
input buffer, unterminated
procedure or interrupted procedure
F Error is shown on the front
panel as it occurs
To a maximum of
36 text characters
DDE Device-Specific Error,
caused by the 5502A because of
some condition, for example,
overrange
R Error is queued to the
remote interface as it occurs
EXE Execution Error, caused by
an element that is inconsistent with
the 5502A functions
S Error causes instrument to
go to Standby
CME Command Error, caused by
incorrect command syntax,
unrecognized header, or parameter
of the incorrect type
D Error causes instrument to
go to the power up condition
(none) Error is given to the
initiator only (i.e., local initiator
or remote initiator)
0 (QYE: ) No Error
1 (DDE:FR ) Error queue overflow
100 (DDE:FR D) Inguard not responding (send)
101 (DDE:FR D) Inguard not responding (recv)
102 (DDE:FR D) Lost sync with inguard
103 (DDE:FR) Invalid guard xing command
104 (DDE:FR D) Hardware relay trip occurred
105 (DDE:FR D) Inguard got impatient