1-3
Introduction
Warning
To prevent possible electrical shock, fire, or personal injury,
read all safety Information before you use the Product.
The 5502A Calibrator (“the Product” or “the Calibrator”) in Figure 1-1 can be configured
to source:
DC voltage from 0 V to ±1020 V.
AC voltage from 1 mV to 1020 V, with output from 10 Hz to 500 kHz.
AC current from 29 μA to 20.5 A, with variable frequency limits.
DC current from 0 to ±20.5 A.
Resistance values from a short circuit to 1100 MΩ.
Capacitance values from 220 pF to 110 mF.
Simulated output for eight types of Resistance Temperature Detectors (RTDs).
Simulated output for 11 types of thermocouples.
0
123
4
5
6
7
8 9
ENTER
M
k
m
V Hz
/
+
F
OPR
STBY
EARTH
TRIG
OUT
SCOPE
ME
NU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
n
p
W
dBm
sec
¡C
POWER
A
MULT
x
DIV
CAL
ZERO
FIELD
EDIT
5502A
CALIBRATOR
gvx001.eps
Figure 1-1. 5502A Multi-Product Calibrator
Calibrator features include:
Automatic meter error calculation, with reference values that you can select.
and keys that change the output value to predetermined values for various
functions.
Programmable entry limits. These limits do not let you go above preset output limits.
Voltage and current that can be output at the same time, to an equivalent of 20.9 kW.
10 MHz reference input and output. Use this to input a high-accuracy 10 MHz
reference to move the frequency accuracy to the 5502A, or to put one or more
Calibrators in the control of a master 5502A.
The power to output two voltages at the same time.
Extended bandwidth mode outputs multiple waveforms down to 0.01 Hz, and sine