SC600 Oscilloscope Calibration Option
SC600 Oscilloscope Calibration Option Specifications 9
9-5
Leveled Sine Wave Specifications
Leveled Sine Wave
Qualities into 50 Ω
Frequency Range
50 kHz (reference) 50 kHz to 100 MHz 100 to 300 MHz 300 to 600 MHz
Amplitude Qualities (for measuring oscilloscope bandwidth)
Range (p-p) 5 mV to 5.5 V
Resolution
<100 mV: 3 digits
100 mV: 4 digits
Adjustment Range continuously adjustable
1-Year Absolute
Uncertainty, tcal ±5 °C
±(2 % of output
+ 300 μV)
±(3.5 % of output
+ 300 μV)
±(4 % of output
+ 300 μV)
±(6 % of output
+ 300 μV)
Flatness (relative to
50 kHz)
not applicable
±(1.5 % of output
+ 100 μV)
±(2 % of output
+ 100 μV)
±(4 % of output
+ 100 μV)
Short-Term Amplitude
Stability
1 %
[1]
Frequency Qualities
Resolution 1 kHz 10 kHz
1-Year Absolute
Uncertainty, tcal ±5 °C
±2.5 ppm
[2]
Distortion Qualities
2nd Harmonic
-33 dBc
3rd and Higher Harmonics
-38 dBc
[1] Within 1 hour after reference amplitude setting, provided temperature varies no more than ±5 °C.
[2] With REF CLK set to ext, the frequency uncertainty of the Leveled Sine Wave is the uncertainty of the external 10 MHz clock
±0.3 Hz/gate time.
Time Marker Specifications
Time Maker into 50 Ω
5 s to 50 ms
20 ms to
100 ns
50 to 20 ns 10 ns 5 to 2 ns
1-Year Absolute
Uncertainty at Cardinal
Points, tcal ±5 °C
±(25 + t *1000)
ppm
[1]
±2.5 ppm ±2.5 ppm ±2.5 ppm ±2.5 ppm
Wave Shape spike or square
spike, square,
or 20 %-pulse
spike or square square or sine sine
Typical Output Level >1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
Typical Jitter (rms) <10 ppm <1 ppm <1 ppm <1 ppm <1 ppm
Sequence 5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
Adjustment Range
[3]
At least ±10 % around each sequence value indicated above.
Amplitude Resolution 4 digits
[1] t is the time in seconds.
[2] Typical rise time of square wave and 20 %-pulse (20 % duty cycle pulse) is < 1.5 ns.
[3] Time marker uncertainty is ±50 ppm away from the cardinal points.
Wave Generator Specifications
Wave Generator Qualities
Square Wave, Sine Wave, and Triangle Wave
into 50 Ω or 1 MΩ
Amplitude
Range
into 1 MΩ: 1.8 mV to 55 V p-p
into 50 Ω: 1.8 mV to 2.5 V p-p
1-Year Absolute Uncertainty, tcal ±5 °C,
10 Hz to 10 kHz
±(3 % of p-p output + 100 µV)
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Typical DC Offset Range
0 to ± (40 % of p-p amplitude)
[1]
Frequency
Range 10 Hz to 100 kHz
Resolution 4 or 5 digits depending upon frequency
1-Year Absolute Uncertainty, tcal ±5 °C ±(25 ppm + 15 mHz)
[1] The DC offset plus the wave signal must not exceed 30 V rms.