Introduction and Specifications
Additional Specifications 1
1-27
AC Current (Non-Sine Wave)
Triangle Wave &
Truncated Sine Wave
Range p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
0.047 to 0.92999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
0.93 to 9.29999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
9.3 to 92.9999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
93 to 929.999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.5
1 to 10 kHz 10 + 2
0.93 to 8.49999 A
[2]
10 to 45 Hz 0.5 + 1.0
Six digits
45 Hz to 1 kHz 0.5 + 0.5
1 to 10 kHz 10 + 2
8.5 to 57 A
[2]
45 to 500 Hz 0.5 + 0.5
500 Hz to 1 kHz 1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.