10-8
Start the Oscilloscope Calibration Option
Push to start the Oscilloscope Calibration Option. The Control Display opens the
volt menu, shown below, which contains options to calibrate the vertical gain on your
oscilloscope. This is the first of five calibration menus, which you can scroll through.
Push the softkey below MODE. Each menu is given in this chapter, see below.
gl021i.eps
The Output Signal
The location of the output signal is shown on the Control Display (the display on the
right side). If your 5502A is connected, but the output does not show on the oscilloscope,
you can have the 5502A in standby mode.
The settings for the output signal are shown in the Output Display (the display on the left
side). The subsequent example shows the default settings for volt mode, which are set
when you start the Oscilloscope Calibration Option, see below.
gl022i.eps
If STBY is shown, push the key. The Output Display will show OPR and the
output will show on the oscilloscope.