Introduction and Specifications
Detailed Specifications 1
1-19
Temperature Calibration (RTD)
RTD Type
Range °C
[1]
Absolute Uncertainty tcal
±5 °C ± °C
[2]
RTD Type
Range °C
[1]
Absolute Uncertainty
tcal ±5 °C ± °C
[2]
90 Day 1 Year 90 Day 1 Year
Pt 385,
100 Ω
-200 to -80 0.04 0.05
Pt 385,
500 Ω
-200 to -80 0.03 0.04
-80 to 0 0.05 0.05 -80 to 0 0.04 0.05
0 to 100 0.07 0.07 0 to 100 0.05 0.05
100 to 300 0.08 0.09 100 to 260 0.06 0.06
300 to 400 0.09 0.10 260 to 300 0.07 0.08
400 to 630 0.10 0.12 300 to 400 0.07 0.08
630 to 800 0.21 0.23 400 to 600 0.08 0.09
Pt 3926,
100 Ω
-200 to -80 0.04 0.05 600 to 630 0.09 0.11
-80 to 0 0.05 0.05
Pt 385,
1000 Ω
-200 to -80 0.03 0.03
0 to 100 0.07 0.07 -80 to 0 0.03 0.03
100 to 300 0.08 0.09 0 to 100 0.03 0.04
300 to 400 0.09 0.10 100 to 260 0.04 0.05
400 to 630 0.10 0.12 260 to 300 0.05 0.06
Pt 3916,
100 Ω
-200 to -190 0.25 0.25 300 to 400 0.05 0.07
-190 to -80 0.04 0.04 400 to 600 0.06 0.07
-80 to 0 0.05 0.05 600 to 630 0.22 0.23
0 to 100 0.06 0.06
PtNi 385,
120 Ω
(Ni120)
-80 to 0 0.06 0.08
100 to 260 0.06 0.07 0 to 100 0.07 0.08
260 to 300 0.07 0.08 100 to 260 0.13 0.14
300 to 400 0.08 0.09
Cu 427
10 Ω
[3]
-100 to 260 0.3 0.3
400 to 600 0.08 0.10
600 to 630 0.21 0.23
Pt 385,
200 Ω
-200 to -80 0.03 0.04
-80 to 0 0.03 0.04
0 to 100 0.04 0.04
100 to 260 0.04 0.05
260 to 300 0.11 0.12
300 to 400 0.12 0.13
400 to 600 0.12 0.14
600 to 630 0.14 0.16
[1] Resolution is 0.003 °C
[2] Applies for COMP OFF (to the 5502A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3] Based on MINCO Application Aid No. 18
Phase
1-Year Absolute Uncertainty, tcal ±5 °C, (Δ Φ °)
Frequency (Hz)
10 to 65 Hz 65 to 500 Hz 500 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz 10 to 30 kHz
0.15 ° 0.9 ° 2 ° 6 ° 10 ° 15 °
Note