Prepare for Operation
Select Line Frequency 2
2-5
Select Line Frequency
The calibrator is shipped from the factory for nominal operation at 60 Hz line frequency.
If you use 50 Hz line voltage, re-configure the Calibrator for optimal performance at
50 Hz. To do this:
1. From the front panel, go into SETUP, INSTMT SETUP, OTHER SETUP.
2. Push the softkey below MAINS to change the selection to 50 Hz.
3. Store the change.
After a correct instrument warmup (on for 30 minutes or longer), you must zero the
complete instrument again. See the section on “Zero the Calibrator” in Chapter 4.
100V/120V
220V/240V
T5.0A 250V
(SB)
T2.5A 250V
(SB)
/63Hz
47Hz
600VA MAX
WARNING:
TO AVOID ELECTRIC SHOCK GROUNDING
CONNECTOR IN POWER CORD MUST BE CONNECTED
FUSE
MAINS SUPPLY
CHASSIS
GROUND
WARNING:
TO AVOID PHYSICAL INJURY, INSURE THA
IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRU
CAUTION
FOR FIRE PROTECTION REPLACE O
WITH A 250V FUSE OF INDICATED RATING
0V
(SB)
120
120
240
Changing Line Fuse
Changing Line
Voltage
Line Voltage
Indicator
gjh004.eps
Figure 2-1. Access the Fuse and Select Line Voltage