5502A
Operators Manual
10-24
4. Do this procedure again for all time marker values recommended for your
oscilloscope. Do this again for digital and analog mode as necessary. It can be
necessary to change some oscilloscopes magnification while you calibrate in analog
mode.
5. Remove the signal with .
Test the Trigger
The oscilloscope can trigger on different waveforms. This can be verified with the wave
generator. When the wave generator is used, a square, sine, or triangle wave is
transmitted and the output impedance of the wave, offset, and voltage can be changed to
examine the trigger function at different levels.
Note
The wave generator must not be used to verify the accuracy of your
oscilloscope. The question mark in the Output Display shows the amplitude
values are not sufficiently accurate for this function.
The wave generator is available through the Wavegen menu, shown below. To access this
menu, push the softkey below MODE until “wavegen” shows, see below.
gl035i.eps