Fluke 5502A - Test RS-232 UUT Port Operation with a Terminal

Fluke 5502A
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5502A
Operators Manual
5-18
UUT
5502A Calibrator
Controller
SERIAL 2
TO UUT
Port
COM Port
RS-232
Port
Null Modem Cable
Modem Cable
gjh045.eps
Figure 5-5. Testing the RS-232 UUT Port with RS-232 Host Port
Terminal This procedure uses the Terminal accessory supplied with Windows (or equal)
to verify RS-232 UUT port operation.
Visual Basic This procedure uses Visual Basic (see Appendix D) to verify RS-232 Host
port and RS-232 UUT port operation.
Test RS-232 UUT Port Operation with a Terminal
Complete the subsequent procedure to verify RS-232 UUT port operation with the
RS-232 Host port with the Windows Terminal accessory (or equal).
1. Complete “RS-232 UUT Port Setup Procedure” to the Calibrator RS-232 UUT port
to the same parameters of the UUT RS-232 port.
2. Complete “Test RS-232 Host Port Operation with a Terminal” to set up the
Calibrator RS-232 Host port to the same parameters of the PC COM port. After Step
9, go to this procedure and continue to Step 3 below.
3. On the Terminal screen, type UUT_SEND “<uut command>“ where <uut command>
is the command you selected for the UUT response, then push <Enter>. See the UUT
result. For example, to send the command REMS to a UUT, use UUT_SEND
“REMS\n” and push <Enter>.
Note that \n is used, which shows a Carriage Return (CR) as the end-of-line
character. Other characters include \r (Line Feed), \t (Tab), \b (Backspace) and
\f (Form Feed). If an end-of-line character is necessary for the UUT commands,
select one or more of the above.
The characters UUT_SEND “<uut command>“ show as they are recorded. If they
did not show on the screen, the RS-232 interface between the PC and Calibrator Host
port is not in operation. Read the “RS-232 Host Port Setup Procedure” and correct
the problem.
4. If the UUT command does not do what it must, see step 3 of the “RS-232 UUT Port
Setup Procedure” to verify the RS-232 UUT port parameters. Also examine the cable
for UUT connection was a modem (not null modem) cable. Make sure the command
was recorded correctly and had the correct end-of-line character(s), if necessary.