SC600 Oscilloscope Calibration Option
Oscilloscope Connections 9
9-7
Overload Measurement Specifications
Source Voltage
Typical ‘On’ Current
Indication
Typical ‘Off’ Current Indication
Maximum Time Limit DC or AC
(1 kHz)
5 to 9 V 100 to 180 mA 10 mA Setable 1 s to 60 s
Oscilloscope Connections
With the cable supplied with the SC600 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 9-1).
To use the external trigger, connect the TRIG output on the Calibrator to the external
trigger connection on your oscilloscope. To use the external trigger and see its signal with
the calibration signal, connect the TRIG output to A different channel. See your
oscilloscope manual for instructions to connect and see an external trigger.
5502A
CALIBRATOR
gvx036.eps
Figure 9-1. Oscilloscope Connection: Channel and External Trigger
Start the SC600 Option
Push (LED on) to select the SC600 Option. The SCOPE menu, shown below,
shows in the Control Display. You can push the first four softkeys to go directly to the
VOLT, EDGE, LEVSINE, and MARKER calibration menus. Push the last softkey to go
to the OTHER menu (also shown below), this lets you access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Push to return to the SCOPE menu from the OTHER menu. This chapter
tells you about each of these menus.