Front-Panel Operation
Set Output 4
4-25
9. Push . The Calibrator erases your entry from the Control Display and copies
it into the Output Display (below is typical).
nn081f.eps
10. Push to activate the Calibrator output.
gvx321f.eps
LCOMP turns inductive compensation on and off. Inductive compensation is
available for frequencies to a maximum of 1 kHz at outputs to a maximum of
239.999 mA, and for frequencies to a maximum of 440 Hz more than 239.999 mA.
OUTPUT shows if the output is on the AUX or 20A terminals. Outputs 3 A or above
are always on the 20A terminals.
WAVE (waveform) selects one of four different types of waveforms: sine wave,
triangle wave, square wave, and truncated sine wave. (See “Waveform Types” later
in this chapter for more information). When a non-sinusoidal waveform is selected,
the Output Display changes the RMS indication to p-p (P
P
).
Set DC Power Output
Note
Tie the terminals NORMAL LO and AUX LO together at the UUT or at the
Calibrator, with the “LO”s softkey selection “tied.”
The Calibrator makes a dc power output by sourcing a dc voltage on the NORMAL
outputs and a dc current on the AUX outputs. Complete the subsequent procedure to set a
dc power output. If you make an entry error, push one or more times to clear the
display, then reenter the value.