Introduction and Specifications
Detailed Specifications 1
1-11
DC Current
Range
Absolute Uncertainty, tcal ±5 °C
±(% of output +μA)
Resolution
Max Compliance
Voltage V
Max Inductive
Load mH
90 Day 1 Year
0 to 329.999 μA
0.012 + 0.02 0.015 + 0.02 1 nA
10
400
0 to 3.29999 mA
0.010 + 0.05 0.010 + 0.05
0.01 μA
10
0 to 32.9999 mA
0.008 + 0.25 0.010 + 0.25
0.1 μA
7
0 to 329.999 mA
0.008 + 3.3 0.010 + 2.5
1 μA
7
0 to 1.09999 A
0.023 + 44 0.038 + 44
10 μA
6
1.1 to 2.99999 A 0.030 + 44 0.038 + 44
10 μA
6
0 to 10.9999 A
(20 A Range)
0.038 + 500 0.060 + 500
100 μA
4
11 to 20.5 A
[1]
0.080 + 750
[2]
0.10 + 750
[2]
100 μA
4
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1-3. The current may be provided
Formula 60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the
output current in amperes. For example, 17 A, at 23 °C could be provided for 60-23-17 = 20 minutes each hour. When the 5502A
is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those
conditions, the allowable "on" time indicated by the formula and Figure 1-3 is achieved only after the 5502A is outputting currents
<5 A for the "off" period first.
[2] Floor specification is 1500 μA within 30 seconds of selecting operate. For operating times >30 seconds, the floor specification is
750 μA.
Range
Noise
Bandwidth 0.1 Hz to 10 Hz p-p Bandwidth 10 Hz to 10 kHz rms
0 to 329.999 μA
2 nA 20 nA
0 to 3.29999 mA 20 nA 200 nA
0 to 32.9999 mA 200 nA
2.0 μA
0 to 329.999 mA 2000 nA
20 μA
0 to 2.99999 A
20 μA
1 mA
0 to 20.5 A
200 μA
10 mA
Ambient
0 °C
20 °C
10 °C
30 °C
40 °C
Current (Amps)
Minutes per Hour
Duty Cycle (%)
0%
10%
20%
30%
40%
60%
70%
80%
50%
5
10
15
20
25
30
35
40
45
50
0
11 12 13 14 15 16 17 18 19 20
Figure 1-3. Allowable Duration of Current >11 A