5502A
Operators Manual
1-14
AC Voltage (Sine Wave) (cont.)
AUX (Auxiliary Output) [dual output mode only]
Range Frequency
[1]
Absolute Uncertainty, tcal
±5 °C ±(% of output + μV)
Resolution Max Burden
Max Distortion and
Noise 10 Hz to
5 MHz Bandwidth
±(% of output +
floor)
90 Day 1 Year
1.0 to
329.999 mV
10 to 20 Hz
0.15 + 370 0.20 + 370
1 μV
5 mA
0.20 + 200 μV
20 to 45 Hz
0.08 + 370 0.10 + 370 0.06 + 200 μV
45 to 1 kHz
0.08 + 370 0.10 + 370 0.08 + 200 μV
1 to 5 kHz
0.15 + 450 0.20 + 450 0.30 + 200 μV
5 to 10 kHz
0.30 + 450 0.40 + 450 0.60 + 200 μV
10 to 30 kHz
4.00 + 900 5.00 + 900 1.00 + 200 μV
0.33 to
3.29999 V
10 to 20 Hz
0.15 + 450 0.20 + 450
10 μV
5 mA
0.20 + 200 μV
20 to 45 Hz
0.08 + 450 0.10 + 450 0.06 + 200 μV
45 to 1 kHz
0.07 + 450 0.09 + 450 0.08 + 200 μV
1 to 5 kHz
0.15 + 1400 0.20 + 1400 0.30 + 200 μV
5 to 10 kHz
0.30 + 1400 0.40 + 1400 0.60 + 200 μV
10 to 30 kHz
4.00 + 2800 5.00 + 2800 1.00 + 200 μV
3.3 to 5 V
10 to 20 Hz
0.15 + 450 0.20 + 450
100 μV
5 mA
0.20 + 200 μV
20 to 45 Hz
0.08 + 450 0.10 + 450 0.06 + 200 μV
45 to 1 kHz
0.07 + 450 0.09 + 450 0.08 + 200 μV
1 to 5 kHz
0.15 + 1400 0.20 + 1400 0.30 + 200 μV
5 to 10 kHz
0.30 + 1400 0.40 + 1400 0.60 + 200 μV
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is <5 mΩ for outputs ≥0.33 V. The AUX output resistance is <1 Ω. The maximum
load capacitance is 500 pF, subject to the maximum burden current limits.