10-12
The V/DIV Menu
The V/DIV menu, shown below, sets the number of volts shown by each division on the
oscilloscope. This menu gives alternative methods to change the output amplitude that
can be more convenient for special oscilloscope uses. To access the V/DIV menu, push
V/DIV from the Volt menu, see below.
gl025i.eps
Each item in the V/DIV menu is given below.
V/div Changes the scale of the output display with a change to the number of volts
that are shown by each division. The available settings, shown in the figure above,
are given in 1-2-5 step increments. Push the softkey below UP to increase the volts
per division. Push the softkey below DOWN to decrease the volts per division.
# DIV Specifies the number of divisions that make the p-p value of the waveform.
The value can be adjusted from one to eight divisions. The quantity shown by each
division is shown in the V/div field. Push the softkey below UP to increase the
height of the signal, and push the softkey below DOWN to decrease it.