5502A
Operators Manual
9-8
gjh050.eps
The Output Signal
If you have selected VOLT mode from the SCOPE menu, see the subsequent description.
The Control Displays shows as follows with VOLT mode selected:
gjh051.eps
The location of the output signal is shown on the Control Display (the display on the right
side). If your Calibrator is connected, but the output does not show on the oscilloscope,
you can have the Calibrator in standby mode. The settings for the output signal are
indicated in the Output Display (the display on the left side).
If STBY is shown, push the key. The Output Display will show OPR and the output
should show on the oscilloscope.
Adjust the Output Signal
The Calibrator has several ways to change the settings for the output signal during
calibration. Since oscilloscope calibration requires many adjustments of the output signal,
the three available methods for changing these settings for oscilloscope calibration are
summarized below. These methods provide the means of jumping to a new value or
sweeping through a range of values.