Introduction and Specifications
Additional Specifications 1
1-25
AC Voltage (Non-Sine Wave) (cont.)
Square Wave
Range (p-p)
[1]
Frequency
1-Year Absolute Uncertainty, tcal
±5 °C, ±(% of output + % of range)
[2]
Max Voltage Resolution
Normal Channel (Single Output Mode)
2.9 to 65.999 mV
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
66 to 659.999 mV
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
0.66 to 6.59999 V
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
6.6 to 66.0000 V
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
Auxiliary Output (Dual Output Mode)
29 to 659.999 mV
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
0.66 to 6.59999 V
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
6.6 to 14.0000 V
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
[1] To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.