Introduction and Specifications
Detailed Specifications 1
1-13
AC Voltage (Sine Wave)
Range Frequency
Absolute Uncertainty, tcal
±5 °C ±(% of output + μV)
Resolution Max Burden
Max Distortion and
Noise 10 Hz to
5 MHz Bandwidth
±(% of output +
floor)
90 Day 1 Year
1.0 to
32.999 mV
10 Hz to 45 Hz
0.120 + 20 0.150 + 20
1 μV 65 Ω
0.15 + 90 μV
45 Hz to 10 kHz
0.080 + 20 0.100 + 20 0.035 + 90 μV
10 kHz to 20 kHz
0.120 + 20 0.150 + 20 0.06 + 90 μV
20 kHz to 50 kHz
0.160 + 20 0.200 + 20 0.15 + 90 μV
50 kHz to 100 kHz
0.300 + 33 0.350 + 33 0.25 + 90 μV
100 kHz to 500 kHz
0.750 + 60 1.000 + 60
0.3 + 90 μV
[1]
33 mV to
329.999 mV
10 Hz to 45 Hz
0.042 + 20 0.050 + 20
1 μV 65 Ω
0.15 + 90 μV
45 Hz to 10 kHz
0.029 + 20 0.030 + 20 0.035 + 90 μV
10 kHz to 20 kHz
0.066 + 20 0.070 + 20 0.06 + 90 μV
20 kHz to 50 kHz
0.086 + 40 0.100 + 40 0.15 + 90 μV
50 kHz to 100 kHz
0.173 + 170 0.230 + 170 0.2 + 90 μV
100 kHz to 500 kHz
0.400 + 330 0.500 + 330
0.2 + 90 μV
[1]
0.33 V to
3.29999 V
10 Hz to 45 Hz
0.042 + 60 0.050 + 60
10 μV
10 mA
0.15 + 200 μV
45 Hz to 10 kHz
0.028 + 60 0.030 + 60 0.035 + 200 μV
10 kHz to 20 kHz
0.059 + 60 0.070 + 60 0.06 + 200 μV
20 kHz to 50 kHz
0.083 + 60 0.100 + 60 0.15 + 200 μV
50 kHz to 100 kHz
0.181 + 200 0.230 + 200 0.2 + 200 μV
100 kHz to 500 kHz
0.417 + 900 0.500 + 900
0.2 + 200 μV
[1]
3.3 V to
32.9999 V
10 Hz to 45 Hz 0.042 + 800 0.050 + 800
100 μV
10 mA
0.15 + 2 mV
45 Hz to 10 kHz 0.025 + 600 0.030 + 600 0.035 + 2 mV
10 kHz to 20 kHz 0.064 + 600 0.070 + 600 0.08 + 2 mV
20 kHz to 50 kHz 0.086 + 600 0.100 + 600 0.2 + 2 mV
50 kHz to 100 kHz 0.192 + 2000 0.230 + 2000 0.5 + 2 mV
33 V to
329.999 V
45 Hz to 1 kHz 0.039 + 3000 0.050 + 3000
1 mV
5 mA, except
20 mA for
45 Hz to
65 Hz
0.15 + 10 mV
1 kHz to 10 kHz 0.064 + 9000 0.080 + 9000 0.05 +10 mV
10 kHz to 20 kHz 0.079 + 9000 0.090 + 9000 0.6 + 10 mV
20 kHz to 50 kHz 0.096 + 9000 0.120 + 9000 0.8 + 10 mV
50 kHz to 100 kHz 0.192 + 80000 0.240 + 80000 1 + 10 mV
330 V to
1020 V
45 Hz to 1 kHz 0.042 + 20000 0.050 + 20000
10 mV
2 mA, except
20 mA for 45
to 65 Hz
0.15 + 30 mV
1 kHz to 5 kHz 0.064 + 20000 0.080 + 20000 0.07 + 30 mV
5 kHz to 10 kHz 0.075 + 20000 0.090 + 20000 0.07 + 30 mV
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9 % of output + floor as shown.
Note
Remote sensing is not provided. Output resistance is <5 mΩ for outputs 0.33 V. The AUX output resistance is <1 Ω. The maximum
load capacitance is 500 pF, subject to the maximum burden current limits.

Table of Contents