5502A
Operators Manual
10-28
Verification Tables
At the Fluke Calibration factory, the SC300 Option is verified to its specifications at the
subsequent test points. The verification test points are given here as a guide when
verification is necessary.
Table 10-1. Voltage Function Verification: AC Voltage into a 1 MΩ Load
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year
Spec.
5.0 mV 10 Hz 0.11 mV
5.0 mV 100 Hz 0.11 mV
5.0 mV 1 kHz 0.11 mV
5.0 mV 5 kHz 0.11 mV
5.0 mV 10 kHz 0.11 mV
10.0 mV 10 kHz 0.12 mV
20.0 mV 100 Hz 0.15 mV
20.0 mV 1 kHz 0.15 mV
20.0 mV 10 kHz 0.15 mV
50.0 mV 10 kHz 0.23 mV
89.0 mV 10 Hz 0.32 mV
89.0 mV 10 kHz 0.32 mV
100.0 mV 10 kHz 0.35 mV
200.0 mV 100 Hz 0.60 mV
200.0 mV 1 kHz 0.60 mV
200.0 mV 10 kHz 0.60 mV
500.0 mV 10 kHz 1.35 mV
890.0 mV 10 Hz 2.32 mV
890.0 mV 10 kHz 2.32 mV
1.0 V 100 Hz 2.60 mV
1.0 V 1 kHz 2.60 mV
1.0 V 10 kHz 2.60 mV
2.0 V 10 kHz 5.10 mV
5.0 V 10 Hz 12.60 mV
5.0 V 10 kHz 12.60 mV
10.0 V 10 kHz 25.10 mV
20.0 V 10 kHz 50.10 mV