SC300 Oscilloscope Calibration Option
Verification Tables 10
10-29
Table 10-1. Voltage Function Verification: AC Voltage into a 1 MΩ Load (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
50.0 V 10 Hz 125.10 mV
50.0 V 100 Hz 125.10 mV
50.0 V 1 kHz 125.10 mV
50.0 V 10 kHz 125.10 mV
105.0 V 100 Hz 262.60 mV
105.0 V 1 kHz 262.60 mV
Table 10-2. Voltage Function Verification: AC Voltage into a 50 Ω Load
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
5.0 mV 10 Hz 0.11 mV
5.0 mV 100 Hz 0.11 mV
5.0 mV 1 kHz 0.11 mV
5.0 mV 5 kHz 0.11 mV
5.0 mV 10 kHz 0.11 mV
10.0 mV 100 Hz 0.12 mV
10.0 mV 1 kHz 0.12 mV
10.0 mV 10 kHz 0.12 mV
20.0 mV 10 kHz 0.15 mV
44.9 mV 10 Hz 0.21 mV
44.9 mV 10 kHz 0.21 mV
50.0 mV 10 kHz 0.23 mV
100.0 mV 100 Hz 0.35 mV
100.0 mV 1 kHz 0.35 mV
100.0 mV 10 kHz 0.35 mV
200.0 mV 10 kHz 0.60 mV
449.0 mV 10 Hz 1.22 mV
449.0 mV 10 kHz 1.22 mV
500.0 mV 10 kHz 1.35 mV
1.0 V 100 Hz 2.60 mV
1.0 V 1 kHz 2.60 mV
1.0 V 10 kHz 2.60 mV
2.0 V 10 Hz 5.10 mV