5502A
Operators Manual
10-30
Table 10-2. Voltage Function Verification: AC Voltage into a 50 Ω Load (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
2.0 V 100 Hz 5.10 mV
2.0 V 1 kHz 5.10 mV
2.0 V 5 kHz 5.10 mV
2.0 V 10 kHz 5.10 mV
Table 10-3. Voltage Function Verification: DC Voltage into a 50 Ω Load
Nominal Value
(dc)
Measured Value (dc) Deviation (mV) 1-Year Spec.
0.0 mV 0.10 mV
5.0 mV 0.11 mV
-5.0 mV 0.11 mV
10.0 mV 0.12 mV
-10.0 mV 0.12 mV
22.0 mV 0.15 mV
-22.0 mV 0.15 mV
25.0 mV 0.16 mV
-25.0 mV 0.16 mV
55.0 mV 0.24 mV
-55.0 mV 0.24 mV
100.0 mV 0.35 mV
-100.0 mV 0.35 mV
220.0 mV 0.65 mV
-220.0 mV 0.65 mV
250.0 mV 0.72 mV
-250.0 mV 0.72 mV
550.0 mV 1.47 mV
-550.0 mV 1.47 mV
700.0 mV 1.85 mV
-700.0 mV 1.85 mV
2.2 V 5.60 mV
-2.2 V 5.60 mV