SC300 Oscilloscope Calibration Option
Summary of Commands and Queries 10
10-27
SCOPE?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Shows the current mode of operation for the oscilloscope. Shows OFF if the oscilloscope
is off.
Parameter: (None)
Response: <character> (Shows OFF, VOLT, EDGE, LEVSINE, MARKER, or
WAVEGEN.)
TRIG
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
Programs the trigger output BNC of the oscilloscope.
Parameters: OFF (Turns off the trigger output.)
DIV1 (Turns on the trigger output. Frequency is the same as the
signal at SCOPE output.)
DIV10 (Turns on the trigger output. Frequency is 1/10 of the
signal at SCOPE output.)
DIV100 (Turns on the trigger output. Frequency is 1/100 of the
signal at SCOPE output.)
TRIG?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Shows the trigger output setting of the oscilloscope.
Parameters: (None)
Response: <character> (Shows OFF, DIV1, DIV10, or DIV100.)
OUT_IMP
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Programs the output impedance of the oscilloscope.
Parameters: Z50 (Programs the output impedance of the oscilloscope to
50Ω.)
Z1M (Programs the output impedance of the oscilloscope to
1 MΩ.)
OUT_IMP?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Shows the output impedance setting of the oscilloscope.
Parameters: (None)
Response: <character> (Shows Z50 or Z1M.)