5502A
Operators Manual
10-26
SCOPE
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Programs the oscilloscope calibration option hardware, if installed. The instrument
settings are sensed by this parameter of this command. When in SCOPE mode, use the
OUT command to program new output.
OPER, STBY, *OPC, *OPC?, and *WAI all operate as given in Chapter 6. The condition
of the oscilloscope output while in SCOPE mode is reflected by the bit in the ISR that is
assigned to SETTLED.
The FUNC? query shows SDCV, SACV, LEVSINE, MARKER, EDGE, and WAVEGEN for
the related oscilloscope modes.
Parameters: OFF Turns off the oscilloscope hardware. Programs 0 V,
0 Hz, output at the NORMAL terminals, standby.
VOLT Oscilloscope ac and dc voltage mode.
Programs 20 mV p-p, 1 kHz, output at the
SCOPE BNC, output impedance 1 MΩ, standby if
from OFF or in standby before.
EDGE Oscilloscope Edge mode. Programs 25 mV peak-to-
peak, 1 MHz, output at the SCOPE BNC, standby if
from OFF or in standby before.
LEVSINE Oscilloscope-leveled sine mode. Programs 30 mV
p-p, 50 kHz, output at the SCOPE BNC,
standby if from OFF or in standby before.
MARKER Oscilloscope Marker mode. Programs the period to
1 ms, output at the SCOPE BNC, standby if from
OFF or in standby before.
WAVEGEN Oscilloscope Wavegen mode. Programs 20 mV
p-p, square wave, 1 kHz, no offset, output
impedance 1 MΩ, standby if from OFF or
in standby before.
Example: SCOPE VOLT; OUT -2 V, 0 Hz (dc voltage, -2 V)
SCOPE VOLT; OUT 4 V, 1 kHz (ac voltage, 4 V peak-to-
peak, 1 kHz.)
SCOPE EDGE; OUT 0.5 V, 5 kHz (Edge, 0.5 V p-p,
5 kHz.)
SCOPE LEVSINE; OUT 1 V, 20 kHz (Leveled sine wave, 2 V
p-p, 20 kHz.)
SCOPE MARKER; OUT 2 MS (Marker, period of 2 ms.)
SCOPE WAVEGEN; OUT 1 V, 1 kHz (Wave Generator, 1 V
p-p, 1 kHz.)