5502A
Operators Manual
4-28
Note
At voltage outputs of 100 V and above (nominal), the Calibrator usually
makes a small high-pitched sound.
5. Push a multiplier key, if necessary. For example, push .
6. Push .
7. The Control Display shows the amplitude of your voltage entry. For example,
123.456 mV (see below).
nn084f.eps
8. Push the numeric keys and decimal point key to record the necessary current output
(maximum six numeric keys). For example, 234.567.
9. Push a multiplier key, if necessary. For example, push .
10. Push .
11. The Control Display shows the amplitude of your voltage and current entries. For
example, 123.456 mV and 234.567 mA (see below).
nn085f.eps
12. Push the numeric keys and decimal point key to record the necessary frequency
output (maximum five numeric keys). Push a multiplier key, if necessary. For
example, push the kilo multiplier key . Then push the key. For example,
1.1234 kHz.
13. The Control Display shows your entries. For example, 123.456 mV and 234.567 mA
at 1.1234 kHz (see below).
nn086f.eps