5502A
Operators Manual
4-60
5502A
CALIBRATOR
87
TRUE RMS MULTIMETER
MIN MAX
RANGE HOLD
H
REL Hz
mA
A
A
m
mV
V
V
OFF
A
COM
V
mA
m
A
PEAK MIN MAX
gvx027.eps
Figure 4-17. Cable Connections to Verify an 80 Series High Amps Function
c. Set the Calibrator output to 3.5 A at 0 Hz and push . Make sure the error is
in specification limits.
d. Do this step again with 0 A at 0 Hz. Make sure the error is in specification limits.
e. Push on the Calibrator and push the blue key on the DMM to change to ac
measurements.
f. Set the Calibrator output to 3.5 A at 60 Hz and push . Make sure the error is
in specification limits.
g. Do this step again with the Calibrator set to:
AC Current Frequency
3.5 A 1.0 kHz
10.0 mA 60 Hz
10.0 mA 1.0 kHz