5502A
Operators Manual
4-62
Table 4-4. Watts Performance, Text Screen
Calibrator Outputs Performance Limits
Normal
V ac @
60 Hz
Phase
in DEG.
AUX
mV ac @
60 Hz
W/KW VA/KVA
VAR/KVAR
Model 41
Only
Phase
Harmonics
Screen
MIN MAX MIN MAX MIN MAX MIN MAX
5.0 V 0.0 30.0 mV 145 156 145 156 0 4 -2 2
8.0 V 0.0 30.0 mV 234 246 234 246 0 4 -2 2
100.0 V 157.0 150.0 mV -14.3k -13.3k 14.5k 15.6k 5.4k 6.3k 155 159
100.0 V 157.0 360.0 mV -37k -29k 32k 40k 10k 18k 155 159
10.0 V 46.0 1.40 V 9.2 10.2 13.5 14.5 9.6 10.6 44 48
100.0 V 46.0 1.40 V 92 102 135 145 96 106 44 48
1. Connect the Calibrator to the Model 41 as shown in Figure 4-18.
Note
Voltage is connected to the Model 41 amps input to simulate current clamp
operation (1 mV = 1 A).
5502A
CALIBRATOR
POWER HARMONICS
ANALYZER
41
®
gvx028.eps
Figure 4-18. Cable Connections to Verify a 40 Series Watts Function